CT Angio Head Select your scanner model below to view the full CTA Head IVcon (IMG14294) protocol for that scanner. Aquillion One CTA Head IVcon — IMG14294 Machine Model: Aquillion One SNIPs Generated: 2021-05-10 14:25:45 Contrast IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines for other access lines. IV Saline 2 ml/sec using 20 g PIV Minimum Oral Contrast NA Water NA Contrast amount determined by weight and renal function. Scan Parameters Special Instructions: IOML perpendicular to table top Scanogram: Lateral / AP Landmark: Hyoid BT Bolus Tracking A Arterial B Venous Scan Start Hyoid Skull Base Skull Base Scan Stop Hyoid Skull Vertex Skull Vertex Patient Position Supine Supine Supine Scan Direction caudal/cranial caudal/cranial caudal/cranial Scan Mode S&V Helical Helical Scan Range If Volume NA NA NA Scan Delay 10 seconds Bolus Track Carotids manually 2 minutes Bolus Tracking Bolus Track Carotids Manually Bolus Track Carotids Manually NA Reconstruction 1 Thickness x Spacing .5 mm x .3 mm 1 mm x 1 mm Algorithm CTA Brain Brain FOV 22 cm 22 cm Reconstruction 2 Thickness x Spacing MIP 10 x 5 Algorithm CTA Brain FOV Axial + Cor + Sag FC 2 43 23 kVp 120 kVp 120 kVp 120 kVp Scan FOV 24 cm 24 cm 24 cm Pitch NA Detail Detail mA 50 mA 250 mA 250 mA Rotation Time .5 sec .5 sec .75 sec Detector Rows 1 mm x 4 mm 0.5 mm x 64 mm 0.5 mm x 32 mm Indications: No indications found. . Scanogram / Scan Range Diagram [Placeholder: scanogram diagram showing BT, A and B scan ranges] Reconstruction Reference CTA Head MIP 10 x 5 — Axial / Coronal / Sagittal [Placeholder: MIP reconstruction plane reference image] Indications: No indications found. . Canon AQ Prime CTA Head IVcon — IMG14294 Machine Model: Canon AQ Prime SNIPs Generated: 2021-05-10 14:13:17 Contrast IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines for other access lines. IV Saline 2 ml/sec using 20 g PIV Minimum Oral Contrast NA Water NA Contrast amount determined by weight and renal function. Scan Parameters Special Instructions: IOML perpendicular to table top Scanogram: Lateral / AP Landmark: Hyoid BT Bolus Tracking A Arterial B Venous Scan Start Hyoid Skull Base Skull Base Scan Stop Hyoid Skull Vertex Skull Vertex Patient Position Supine Supine Supine Scan Direction caudal/cranial caudal/cranial caudal/cranial Scan Mode S&V Helical Helical Scan Range If Volume NA NA NA Scan Delay 10 seconds NA 2 minutes Bolus Tracking Bolus Track Carotids Manually Bolus Track Carotids Manually NA Reconstruction 1 Thickness x Spacing .5 mm x .3 mm 1 mm x 1 mm Algorithm CTA Brain Brain FOV 22 cm 22 cm Reconstruction 2 Thickness x Spacing MIP 10 x 5 Algorithm CTA Brain FOV Axial + Cor + Sag FC 2 43 23 kVp 120 kVp 120 kVp 120 kVp Scan FOV 24 cm 24 cm 24 cm Pitch NA Detail Detail mA 50 mA 300 mA 300 mA Rotation Time .5 sec .5 sec .5 sec Detector Rows 1 mm x 4 mm 0.5 mm x 40 mm 0.5 mm x 40 mm Indications: No indications found. . Scanogram / Scan Range Diagram [Placeholder: scanogram diagram showing BT, A and B scan ranges] Reconstruction Reference CTA Head MIP 10 x 5 — Axial / Coronal / Sagittal [Placeholder: MIP reconstruction plane reference image] Indications: No indications found. . Canon AQ 64 CTA Head IVcon — IMG14294 Machine Model: Canon AQ 64 SNIPs Generated: 2021-05-10 14:13:57 Contrast IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines for other access lines. IV Saline 2 ml/sec using 20 g PIV Minimum Oral Contrast NA Water NA Contrast amount determined by weight and renal function. Scan Parameters Special Instructions: IOML perpendicular to table top Scanogram: Lateral / AP Landmark: Hyoid BT Bolus Tracking A Arterial B Venous Scan Start Hyoid Skull Base Skull Base Scan Stop Hyoid Skull Vertex Skull Vertex Patient Position Supine Supine Supine Scan Direction caudal/cranial caudal/cranial caudal/cranial Scan Mode S&V Helical Helical Scan Range If Volume NA NA NA Scan Delay 10 seconds NA 2 minutes Bolus Tracking Bolus Track Carotids Manually Bolus Track Carotids Manually NA Reconstruction 1 Thickness x Spacing .5 mm x .3 mm 1 mm x 1 mm Algorithm CTA Brain Brain FOV 22 cm 22 cm Reconstruction 2 Thickness x Spacing MIP 10 x 5 Algorithm CTA Brain FOV Axial + Cor + Sag FC 2 43 23 kVp 120 kVp 120 kVp 120 kVp Scan FOV 24 cm 24 cm 24 cm Pitch NA Detail Detail mA 50 mA 300 mA 300 mA Rotation Time .5 sec .5 sec .5 sec Detector Rows 1 mm x 4 mm 0.5 mm x 64 mm 0.5 mm x 64 mm Indications: No indications found. . Scanogram / Scan Range Diagram [Placeholder: scanogram diagram showing BT, A and B scan ranges] Reconstruction Reference CTA Head MIP 10 x 5 — Axial / Coronal / Sagittal [Placeholder: MIP reconstruction plane reference image] Indications: No indications found. . Canon AQ 16 CTA Head IVcon — IMG14294 Machine Model: Canon AQ 16 SNIPs Generated: 2021-05-10 14:14:07 Contrast IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines for other access lines. IV Saline 2 ml/sec using 20 g PIV Minimum Oral Contrast NA Water NA Contrast amount determined by weight and renal function. Scan Parameters Special Instructions: IOML perpendicular to table top Scanogram: Lateral / AP Landmark: Hyoid BT Bolus Tracking A Arterial B Venous Scan Start Hyoid Skull Base Skull Base Scan Stop Hyoid Skull Vertex Skull Vertex Patient Position Supine Supine Supine Scan Direction caudal/cranial caudal/cranial caudal/cranial Scan Mode S&V Helical Helical Scan Range If Volume NA NA NA Scan Delay 10 seconds NA 2 minutes Bolus Tracking Bolus Track Carotids Manually Bolus Track Carotids Manually NA Reconstruction 1 Thickness x Spacing .5 mm x .3 mm 1 mm x 1 mm Algorithm CTA Brain Brain FOV 22 cm 22 cm Reconstruction 2 Thickness x Spacing MIP 10 x 5 Algorithm CTA Brain FOV Axial + Cor + Sag FC 2 43 23 kVp 120 kVp 120 kVp 120 kVp Scan FOV 24 cm 24 cm 24 cm Pitch NA Detail Detail mA 50 mA 300 mA 300 mA Rotation Time .5 sec .5 sec .5 sec Detector Rows 0.5 mm x 4 mm 0.5 mm x 16 mm 0.5 mm x 16 mm Indications: No indications found. . Scanogram / Scan Range Diagram [Placeholder: scanogram diagram showing BT, A and B scan ranges] Reconstruction Reference CTA Head MIP 10 x 5 — Axial / Coronal / Sagittal [Placeholder: MIP reconstruction plane reference image] Indications: No indications found. . Canon Genesis 640 CTA Head IVcon — IMG14294 Machine Model: Canon Genesis 640 SNIPs Generated: 2021-05-10 14:13:32 Contrast IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines for other access lines. IV Saline 2 ml/sec using 20 g PIV Minimum Oral Contrast NA Water NA Contrast amount determined by weight and renal function. Scan Parameters Special Instructions: IOML perpendicular to table top Scanogram: Lateral / AP Landmark: Hyoid BT Bolus Tracking A Arterial B Venous Scan Start Hyoid Skull Base Skull Base Scan Stop Hyoid Skull Vertex Skull Vertex Patient Position Supine Supine Supine Scan Direction caudal/cranial caudal/cranial caudal/cranial Scan Mode S&V Helical Helical Scan Range If Volume NA NA NA Scan Delay 10 seconds NA 2 minutes Bolus Tracking Bolus Track Carotids Manually Bolus Track Carotids Manually NA Reconstruction 1 Thickness x Spacing .5 mm x .3 mm 1 mm x 1 mm Algorithm CTA Brain Brain FOV 22 cm 22 cm Reconstruction 2 Thickness x Spacing MIP 10 x 5 Algorithm CTA Brain FOV Axial + Cor + Sag FC 2 43 23 kVp 120 kVp 120 kVp 120 kVp Scan FOV 24 cm 24 cm 24 cm Pitch NA Detail Detail mA 50 mA 400 mA 400 mA Rotation Time .5 sec .5 sec .5 sec Detector Rows 0.5 mm x 8 mm 0.5 mm x 80 mm 0.5 mm x 80 mm Indications: No indications found. . Scanogram / Scan Range Diagram [Placeholder: scanogram diagram showing BT, A and B scan ranges] Reconstruction Reference CTA Head MIP 10 x 5 — Axial / Coronal / Sagittal [Placeholder: MIP reconstruction plane reference image] Indications: No indications found. . Canon Vision 640 CTA Head IVcon — IMG14294 Machine Model: Canon Vision 640 SNIPs Generated: 2021-05-10 14:13:43 Contrast IV Contrast 4 ml/sec using 20 g PIV Minimum See IV Guidelines for other access lines. IV Saline 2 ml/sec using 20 g PIV Minimum Oral Contrast NA Water NA Contrast amount determined by weight and renal function. Scan Parameters Special Instructions: IOML perpendicular to table top Scanogram: Lateral / AP Landmark: Hyoid BT Bolus Tracking A Arterial B Venous Scan Start Hyoid Skull Base Skull Base Scan Stop Hyoid Skull Vertex Skull Vertex Patient Position Supine Supine Supine Scan Direction caudal/cranial caudal/cranial caudal/cranial Scan Mode S&V Helical Helical Scan Range If Volume NA NA NA Scan Delay 10 seconds NA 2 minutes Bolus Tracking Bolus Track Carotids Manually Bolus Track Carotids Manually NA Reconstruction 1 Thickness x Spacing .5 mm x .3 mm 1 mm x 1 mm Algorithm CTA Brain Brain FOV 22 cm 22 cm Reconstruction 2 Thickness x Spacing MIP 10 x 5 Algorithm CTA Brain FOV Axial + Cor + Sag FC 1 43 64 kVp 120 kVp 120 kVp 120 kVp Scan FOV 24 cm 24 cm 24 cm Pitch NA Standard Standard mA 50 mA 300 mA 300 mA Rotation Time .5 sec .5 sec .5 sec Detector Rows 1 mm x 4 mm 0.5 mm x 64 mm 0.5 mm x 64 mm Indications: No indications found. . Scanogram / Scan Range Diagram [Placeholder: scanogram diagram showing BT, A and B scan ranges] Reconstruction Reference CTA Head MIP 10 x 5 — Axial / Coronal / Sagittal [Placeholder: MIP reconstruction plane reference image] Indications: No indications found. .